CORRELATION OF INTRAOCULAR PRESSURE MEASUREMENTS WITH NON CONTACT TONOMETER AND GOLDMANN APPLANATION TONOMETRY

Abstract

Leya Sara George, Arun Bhatti, Nitin Batra, Rupali Chopra

BACKGROUND
A complete ophthalmologic examination includes intraocular pressure (IOP) measurement, which is a routine procedure and is important for diagnosis and monitoring of glaucoma. IOP measurement is most commonly done using Goldmann Applanation tonometer and Non-Contact tonometer.
MATERIALS AND METHODS
In this study IOP measurements of 500 eyes (glaucomatous and non-glaucomatous) were performed using GAT and NCT on patients visiting the outpatient clinic of Department of Ophthalmology at Christian Medical College and Hospital, Ludhiana. This was a cross sectional and observational study. Comparison of IOP values was done in different IOP ranges. CCT was measured and analysis of its correlation with GAT and NCT was done.
RESULTS
Both methods of tonometry correlated significantly in patients with IOP <24 mm Hg. The mean IOP measured by NCT, was 16.06 ± 5.85 mm Hg and the mean IOP measurement by GAT was 16.61 ± 6.94 mm Hg. Intraocular pressure readings with GAT and NCT positively correlated with CCT.
CONCLUSION
NCT may be useful for screening in clinical settings but borderline high IOP readings should be confirmed with GAT. Our findings, also suggest that CCT is an essential variable to consider in interpreting IOP readings